SDCHNS536 Carbon/Hydrogen/Nitrogen/Sulfur Analyzer
Standard Configuration
Carbon/hydrogen/nitrogen analyzer, sulfur analyzer, PC
CHN analyzer dimension: 680mm*640mm*996mm
CHN analyzer weight: approx. 160kg
Sulfur analyzer dimension: 760mm*434mm*752mm
Sulfur analyzer weight: approx. 80kg
Highlights
1.Powerful and Cost-effective.
①One set of equipment can complete all the tests of four elements of CHNS, and also support individual tests of a single or multiple elements.
②CHN analysis is integrated with automatic sample delivery, experiment, and cyclic testing, which can be completely unattended. 36 samples autoloader with optional expanded capacity for up to 72 samples, and samples can be added/replaced during analysis.
③Sulfur analysis sample loading is by manually, quick and easy to operate, and no complicated mechanical actions.
2.Full Process Control and High Precision
①Sample introduction component adopted air curtain design, it can prevent the air into the furnace. and auxiliary by flow equalize unit to avoid sample splashing when dropping, ensure complete combustion of the sample and the test result is accurate.
②Equipped high-performance and precision IR sensor to analysis the carbon and hydrogen element, and can monitoring the combustion situation in real time, accurately judge the end of the combustion, effectively shorten the test time and improve the test accuracy.
③High sensitivity and ultra-low temperature drift diffusion type thermal conductivity cell which with independent and accurate thermostatic control to ensure the thermal conductivity baseline stable.
④Ultra-low drift infrared cell, just need to carry out one drift correction per day.
3.Short and Reliable Gas Path
Adopted confluence board to integrated the gas path, low dead volume, short gas path and less water vapor residue, accurate and automatic flow control, gas path reliability improved.
4.Humanized Design, Easy to Operate.
①Easy and convenient to check the gas tightness, only need to click on the software.
②The crucible can be replaced from the top, the furnace body is easy to disassemble and assemble, and furnace reagent is easy to replace.
③Optional high-performance touch screen, with multi-functional retractable and foldable holder.
5.Low Operating Cost
High stability IR sensor can be stabilized in short time. Equipped with a helium throttling device, the helium flow is only 0.05L/min before sample introduced and automatic switch to 0.7L/min after sample dropped, more efficiency than competitor and saving operating cost.
Specifications:
Description |
SDCHNS536 |
Measuring Range |
C(0.005%~100%) H(0.01%~50%) N(0.01%~50%) S(0.01%~50%) |
Autoloader |
CHN: 36/72 S: 1 |
Analysis time | ≤5min |
Sample weight |
CHN: 75mg~105mg(recommend:100mg) S: 100mg~500mg (recommend:100mg) |
Power supply |
220VAC±10%, 50/60Hz, 8.5kW |
Repeatability |
C≤0.5% H≤0.15% N≤0.08% S: ≤0.05% (St,d≤1.5%) ≤0.1%(1.50%<St,d≤4%) ≤0.2% (St,d>4%) |
Gas required |
Combustion-supporting: Oxygen, ≥99.5% pure, ≥1MPa Driving: Nitrogen, 99.9% pure, ≥1MPa Carrier: Helium, 99.99% pure, ≥1MPa |
Standard | AS 1038.6.4, ASTM D4239,ASTM D5373, ASTM D5291, ISO 16634, ISO 16948, ISO 29541, ISO 19579:2006,EN 15407,EN ISO 16948:2015-0, GB/T 30728, GB/T 30733, ,GB/T25214-2010,UNI 15104, UNI CEN/TS 15407 |