Sales Contact
0086 731 88112150
0086 731 89864000
0086 731 88134650
Quick Links
Hunan Sundy Science and Technology CO.,LTD
No.558 West Tongzipo Road,
Yuelu Disctrict,
Changsha, Hunan Province, China P.R
Copyright©HunanSundy Science and Technology Co., Ltd. 2017 湘ICP备10204722号 www.300.cn Changsha Backstage
SDTGA6000A Thermogravimetric Analyzer (Proximate Analyzer)
①19 samples can be loaded once + 1 blank sample position for analyzing the moisture, ash and volatile matter.
②Unique sample weighing mode, single sample or batch sample weighing are both available, makes the operation convenient.
③Automatic temperature calibration function is available.
④Flexible software design, test procedure, analysis time, temperature etc. can be customized.
⑤LED displays on furnaces to show real-time balance weight data, it makes the sample weighing more convenient.
⑥Single and dual furnace configurations available.
Description |
SDTGA6000A |
Test index |
Moisture, ash and volatile matter |
Continuous test in batches |
Yes |
Standards |
ASTM D3173/D3174/D3175/DD7582, ISO 17246, GB/T 30732, GB/T 212, ISO 18123/18122/18134, ISO 11722, ISO 1171 |
Precision |
0.02% RSD (1σ, 1g sample) |
Furnace temperature |
Room temperature~1000℃ |
Temperature control precision |
±2℃ |
Temperature resolution |
0.1℃ |
Reproducibility |
Ash: 0.30% (<15.00%); 0.50% (15.00~30.00%); 0.70% (>30.00%)Volatile matter: 0.50% (<20.00%); 1.00% (20.00~40.00%); 1.50% (>40.00%) |
Cooling time required for batches of test |
No |
Furnace Cooling (forced air/N2) |
950 to 500°C<30 min |
Balance Resolution |
0.0001g |
Sample introduction and taking |
Automatic |
Max. Samples |
19 samples + 1 blank sample position |
Quantity of sample / test time: |
Full indexes for 19 samples< 120 min (Quick method) |
Sample weight |
0.5 up to 5g |
Max. power |
4kW |
Interface: |
RS232 |
Furnace Heating |
Resistance Wound |
Temperature Calibration |
Standard thermocouple |
Temp. control precision |
±2℃ (moisture) ±5℃ (ash, volatile matter) |
Gas Flow Rate |
GB:Moisture:Nitrogen,30~60 times/h Ash:Air 60~120 times/h ASTM:Moisture:Nitrogen or Air 2~4(Furnace Volume /min)volatile matter:Nitrogen 2~4(Furnace Volume /min) |
Power requirement |
220 V(±10%),50 / 60Hz |
Gas requirement |
Nitrogen/Oxygen/Compressed Air |
Gas Pressure |
Air 5 to 6 bar (75 to 90 psi) |
Minimum Gas Purity |
Air/Nitrogen/Oxygen 99.5%; dry, oil-free |
Sample weighing |
Under normal temp. |
Test Temperature |
Moisture 105℃,Ash 815℃,Volatile matter 920℃~950℃ Can be customized |
Humidity |
<80% |
Dimensions |
550*590*530mm |
Weight |
70kg |
Operational Control |
Windows®-based software on external PC/Windows7/ WindowsXP |
Advance Calculation |
Yes |
Analysis Report |
Yes |
Statistics Report |
Yes |
System Monitor |
Yes |